PARAMETER |
SPECIFICATION |
Working Condition |
Water input pressure |
0.1Mpa~0.6Mpa, 15P.S.I~60P.S.I |
Water input temperature |
5℃~30℃ |
Working environment temperature |
10℃~30℃ at relative humidity ≦70% |
UF Rate |
Flow range |
0ml/h~4000ml/h |
Resolution ratio |
1ml |
Precision |
±30ml/h |
Blood Pump & Substitution Pump |
Blood pump flow range |
10ml/min~600ml/min (diameter: 8mm or 6mm) |
Substitution pump flow range |
10ml/min~300ml/min (diameter 8mm or 6mm) |
Resolution ratio |
0.1ml |
Precision |
±10ml or 10% of reading |
Heparin Pump |
Syringe size |
20, 30, 50ml |
Flow range |
0ml/h~10ml/h |
Resolution ratio |
0.1ml |
Precision |
±5% |
Monitoring System & Alarm Setup |
Venous pressure |
-180mmHg ~ +600mmHg, ±10mmHg |
Arterial pressure |
-380mmHg ~ +400mmHg, ±10mmHg |
TMP |
-180mmHg ~ +600mmHg, ±20mmHg |
Dialysate temperature |
preset range 34.0℃~39.0℃ |
Dialysate flow |
Less than 800 ml/min (Adjustable) |
Substitution flow range |
0-28 L/H (on line HDF) |
Blood leak detection |
Photo chromic alarm when erythrocyte specific volume is 0.32±0.02 or blood leak volume is equal or more than 1ml per liter of dialysate |
Bubble detection |
Ultrasonic, Alarm when a single air bubble volume is more than 200μl at 200ml/min blood flow |
Conductivity |
Acoustic-optic |
Disinfection/Sanitize |
Hot disinfection |
Time: 30 minutes |
Temperature: about 80℃, at flow rate 500ml/min |
Chemical disinfection |
Time: 30 minutes |
Temperature: about 36℃~50℃, at flow rate 500ml/min |
Chemical disinfection with heat |
Time: 45minutes |
Temperature: about 36℃~80℃, at flow rate 50ml/min |
Rinse |
Time: 10minutes |
Temperature: about 37℃, at flow rate 800ml/min |
Storage Environment |
Storage temperature should be between 5℃~40℃, at relative humidity ≦80% |
Power Supply |
AC220V, 50Hz/60Hz, 10A |
Input power |
1500W |
Back-up battery |
30 minutes |
Size &Weight |
Size |
380mmx400x1380mm (L*W*H) |
Net Weight approx. |
88KG |
Gross Weight approx. |
around 100KG |
Package Size approx |
650×690×1581mm (L x W x H) |
Function |
HDF, on-line BPM, Bi-cart and 2 pcs endotoxin filters |
Optional Function |
Online KT/V, LAN |